Model Number M00542250
Device Problems Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/25/2020
Event Type  malfunction  
Manufacturer Narrative
The complainant was unable to report a valid device lot number; therefore, the manufacture date and expiration date are unknown.(b)(6).(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used during a band ligation procedure performed on (b)(6) 2020.According to the complainant, during the procedure, there was no click sound while rotating the handle and the trip wire was not attached spool track.The procedure was completed with another speedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDBAND SUPERVIEW SUPER 7
Type of Device
LIGATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key11206853
MDR Text Key227949471
Report Number3005099803-2021-00168
Device Sequence Number1
Product Code MND
UDI-Device Identifier08714729201953
UDI-Public08714729201953
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
510K EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00542250
Device Catalogue Number4225
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/30/2020
Initial Date FDA Received01/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-