Model Number BI70002000
Device Problems Computer Operating System Problem (2898); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2021
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that the generator was not initializing ps1 voltage at 4.8v.The ps1 voltage was adjusted according to the advanced service manual.The system passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the imaging system got stuck in initialing please wait.It was noted that before getting stuck the system took a long time to power up.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX  45640
Manufacturer (Section G)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX   45640
Manufacturer Contact
tricha miles
7000 central avenue ne rcw215
minneapolis, MN 55432
7635140379
MDR Report Key11211531
MDR Text Key232917411
Report Number3006544299-2021-00029
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00763000132378
UDI-Public00763000132378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200074
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/19/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-