• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G156
Device Problems High impedance (1291); High Capture Threshold (3266)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2020
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that this system exhibited an out of range pacing impedance measurement greater than 3000 ohms on the right ventricular channel.A review of the stored data identified a small impedance increase over the past few months.System evaluation during a programmer session produced a threshold issue and the out of range measurement was confirmed.A boston scientific technical services consultant discussed the clinical observations with the caller.The patient will continue to be monitored.No adverse patient effect were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11212148
MDR Text Key228162816
Report Number2124215-2021-01251
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534669
UDI-Public00802526534669
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/20/2020
Device Model NumberG156
Device Catalogue NumberG156
Device Lot Number379489
Was Device Available for Evaluation? No
Date Manufacturer Received12/14/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/13/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
-
-