Cooling device failed.Per report, when it was placed, an error message appeared - e113.This is a water temperature notification that it is not being temperature controlled.The device was pulled out of service and tested by biomedical department.They noted another error message -e80.This means the control processor failed to detect a simulated water temperature fault.Biomed contacted the vendor who reported the mixing pump was not working correctly.It was not pushing enough cold water into the lines to cool the patient to the correct temperature.The vendor noted the pump needed to be replaced.The unit was sent to the vendor for repair.Biomed also had another unit with the same issue and it too was sent for repair.Biomed reported the units had appropriate preventative maintenance.Manufacturer has not responded to date for either unit.
|