It was reported that this left ventricular exhibited high-out-of-range pacing impedances measuring greater than 2500 ohms.Additionally, the lead did not sense and has loss of capture.It was noted there was no noise that was observed.The patient is scheduled for a revision procedure and they will encourage the lead to be returned for product analysis.At this time, this lv lead remains in service.No adverse patient effects were reported.
|
It was reported that this left ventricular exhibited high-out-of-range pacing impedances measuring greater than 2500 ohms.Additionally, the lead did not sense and has loss of capture.It was noted there was no noise that was observed.The patient is scheduled for a revision procedure and they will encourage the lead to be returned for product analysis.At this time, this lv lead remains in service.No adverse patient effects were reported.
|