The initial reporter received questionable elecsys ft3 iii results for two patients tested on a cobas 8000 e 801 module serial number (b)(4).The customer reported the ft3 results to a physician who asked for re-measurement of the samples.The samples were sent for an investigation and were tested on a cobas 8000 e 801 module, cobas 8000 e 602 module, cobas e 411 immunoassay analyzer, and a siemens centaur analyzer.Refer to the attachment on the medwatch for all patient data.
|
The customer's calibration, qc, and sample pre-analytic data were requested but not provided.Upon further investigation of patient 1's and patient 2's samples, an interferent against a component of the reagent was not confirmed.Per product labeling, "for diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings." the investigation did not identify a product problem.The cause of the event could not be determined.
|