MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN X4 CRT-D; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number G148
Device Problem Over-Sensing (1438)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2020
Event Type  malfunction  
Manufacturer Narrative
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.Investigation of the available information determined this device exhibited oversensing of noise generated by the minute ventilation (mv)/respiratory sensor that is related to a high impedance condition.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) stored an inappropriate atrial tachy response (atr) due to oversensing on the atrial channel.Technical services discussed programming off the respiratory rate trend sensor to mitigate further oversensing.Subsequently, the atrial lead implanted with this device was replaced.Records indicate the device remains in service without further complication.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11218899
MDR Text Key228400591
Report Number2124215-2021-01538
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/24/2018
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number155909
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/18/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
-
-