Model Number CAT6
Device Problems Stretched (1601); Material Twisted/Bent (2981)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2020
Event Type  malfunction  
Event Description
The patient was undergoing a thrombectomy procedure in the superior mesenteric artery (sma) using an indigo system aspiration catheter 6 (cat6) and neuron max 6f 088 long sheath (neuron max).During the procedure, the physician removed thrombus proximally in the sma using the neuron max.The physician then noticed a bifurcation of the sma that became thrombosed.Subsequently, while inserting the cat6 using the peel away sheath, it was noticed under fluoroscopy that the cat6 became bent between the mid shaft and towards the distal end.Therefore, the cat6 was removed and upon further inspection, the physician confirmed the cat6 was stretched.The procedure was completed by administering tissue plasminogen activator (tpa).There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, an investigation cannot be performed.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 6
Type of Device
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key11219350
MDR Text Key228406037
Report Number3005168196-2021-00153
Device Sequence Number1
Product Code DXE
UDI-Device Identifier00814548016245
UDI-Public00814548016245
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
K160533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/20/2021
Device Model NumberCAT6
Device Catalogue NumberCAT6
Device Lot NumberF84232
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 12/30/2020
Initial Date FDA Received01/22/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age35 YR
-
-