Model Number G447
Device Problems Off-Label Use (1494); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Pocket Erosion (2013); Sepsis (2067); Fluid Discharge (2686); No Code Available (3191)
Event Date 10/23/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.(b)(4).Upon receipt at our post market quality assurance laboratory that the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) was part of a system revision due to infection and erosion with sepsis.Reportedly, the patient had a pocket infection from recent upgrade.The patient noted that the incision was oozing since the upgrade and opened within the last week.The patient was treated with intravenous antibiotics.There were no additional adverse patient effects reported.The crt-d was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESONATE X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key11220787
MDR Text Key241237622
Report Number2124215-2020-28843
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526589539
UDI-Public00802526589539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S436
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date07/29/2022
Device Model NumberG447
Device Catalogue NumberG447
Device Lot Number242363
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/23/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age64 YR
-
-