HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CS ELITE PROCESSING KIT, 225ML Back to Search Results
Model Number CSE-P-225
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/12/2020
Event Type  malfunction  
Manufacturer Narrative
On january 12, 2021 haemonetics manufacturing performed a visual inspection of the received bowl from the cell saver® elite set - 225ml and confirmed blood in the inner core with a crack in the inner core base.Although there was no serious injury or harm, past reporting (1219343-2020-00001-01) indicates this particular malfunction on a similar device has been associated with a reported death event.The investigation of 1219343-2020-00001-01 indicated the inner core of the bowl malfunctioned via a crack but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core cracks that are confirmed by manufacturing due the event of the past report.
 
Event Description
On (b)(6) 2020 haemonetics was notified of a returning error alarm which was observed during a procedure, utilizing the cell saver 5 autologous blood recovery system and cell saver® elite set - 225ml.There was no reported impact to patients' health.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE AUTOTRANSFUSION SYSTEM
Type of Device
CS ELITE PROCESSING KIT, 225ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer (Section G)
HAEMONETICS CORPORATION
125 summer street
boston MA 02110
Manufacturer Contact
janice kiser
125 summer street
boston, MA 02110
MDR Report Key11221940
MDR Text Key241614683
Report Number1219343-2020-00162
Device Sequence Number1
Product Code CAC
UDI-Device Identifier10812747016544
UDI-Public(01)10812747016544(17)230804(10)0820013
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120586
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 01/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/04/2023
Device Model NumberCSE-P-225
Device Lot Number0820013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/12/2021
Date Manufacturer Received01/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/04/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-