Model Number N/A |
Device Problem
Intermittent Loss of Power (4016)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once the investigation is complete, a follow up/final report will be submitted.Telephone number: (b)(6).
|
|
Event Description
|
It was reported that the handpiece works intermittently.There are problems with start up.When the head is touched, the unit starts and works correctly.After several trials (with no blade) the rollers stalls again.Restart when touched.Initially issue noted during surgery, verified and confirmed in technical dept.No adverse consequences reported.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the motor speed was high.The motor was replaced and resolved the reported issue.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event cannot be confirmed.
|
|
Event Description
|
No additional event information.
|
|