Model Number IPN001112 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that the intra-aortic balloon pump (iabp) had a 20-minute alarm and then shut down shortly thereafter during patient transport.The iabp was not swapped out.There was no report of patient complications, serious injury or death.
|
|
Manufacturer Narrative
|
(b)(4).Teleflex received the device for investigation.The reported complaint of "dead battery" is not confirmed; however, the pump shut off approximately 50 minutes when operating on battery power after a full charge during.The battery failed battery load test.A definitive root cause could not be determined but a potential cause of the short battery life is a result of maintaining of the battery.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
|
|
Event Description
|
It was reported that the intra-aortic balloon pump (iabp) had a 20-minute alarm and then shut down shortly thereafter during patient transport.The iabp was not swapped out.There was no report of patient complications, serious injury or death.
|
|