• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2020
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of the autopulse platform (serial #(b)(4)) power off by itself was not confirmed during functional testing.The autopulse platform powers on and doesn't power off by itself.The initial functional testing failed, the autopulse platform displayed user advisory "(ua)41" (patient temperature sensor failure) error message upon powering on, unrelated to the reported complaint.The root cause of ua41 error was due to a defective temperature sensor likely due to defective component.The temperature sensor was replaced to remedy ua41.No physical damaged observed during visual inspection.During archive data review, observed multiple ua41 (patient temperature sensor failure) error messages in the archive, unrelated to the reported complaint.After service repair completion, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test batteries until discharged without any fault or error.The autopulse passed all functional testing.
 
Event Description
During reanimation training on a manikin, customer reported that the autopulse platform (serial #(b)(4)) power off by itself.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key11228303
MDR Text Key229140981
Report Number3010617000-2021-00042
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000536
UDI-Public00849111000536
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-08
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/28/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/29/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-