MEDTRONIC HEART VALVES DIVISION ENVEO PRO DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
|
Back to Search Results |
|
Model Number ENVPRO-16 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/05/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, during the deployment, the delivery catheter system (dcs) handle came apart and could not be reassembled.The valve was withdrawn from the patient.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Product analysis: the device was received with the handle components detached from the delivery catheter system (dcs), the capsule fully opened, and with the end cap/screw gear snap fit connected.Delamination was observed over the nitinol reinforcing frame along the proximal end of the capsule.The trigger moved to fully advanced and retracted positions and locked in place when released.The tip-retrieval mechanism was intact.A kink was observed along the distal section of the inner member near the nose cone, and along the proximal end, close to the strain relief.One tab was observed to have detached from the male actuator component and a non-smooth transition was visible were this tab detached.A different tab observed to be hinged inward.The horizontal face of the actuator body appeared flat.The reported actuator separation was confirmed in analysis.Conclusion: a device history record (dhr) review was performed on the delivery catheter system (dcs) and there were no correlations / issues identified regarding manufacturing.The device was manufactured per approved and released manufacturing processes and the device met all applicable manufacturing specifications prior to release for distribution.Investigation concluded that the primary root cause of the fractured and disengaged actuator halves from the mating half were due to increased stress on the actuator tabs.Updated data: h6 method, result, and conclusion codes medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|