C.R. BARD, INC. (COVINGTON) -1018233 BARDIA FOLEY CATHETER SILICONE COATED Back to Search Results
Model Number 123520A
Device Problem Burst Container or Vessel (1074)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/31/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the catheter balloon was found to be broken before placing.Customer replaced with new products.
 
Event Description
It was reported that the catheter balloon was found to be broken before placing.Customer replaced with new products.
 
Manufacturer Narrative
The reported event was confirmed and cause unknown.The reported failure was able to be reproduced.The product was used for urological care.Based on the attached photo, it was observed that the balloon catheter could have burst causing leak.However, the potential root cause for this failure mode could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/ operator error.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "uses: to be used as the drainage of urine from the bladder for the children and adults.Warnings: - do not use petroleum substrate lubricants with the latex-based urethral cathethers such as petroleum jelly and label liquid paraffin which will damage latex and cause burst balloon.- do not aspirate urine through the drainage funnel wall.- single use only.Do not re-sterlize.- for urological use only." corrections: d,h h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : photo sample received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA FOLEY CATHETER SILICONE COATED
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key11229332
MDR Text Key228995172
Report Number1018233-2021-00122
Device Sequence Number1
Product Code EZC
UDI-Device Identifier00801741095115
UDI-Public(01)00801741095115
Combination Product (y/n)N
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number123520A
Device Catalogue Number123520A
Device Lot NumberMYEN1840
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/07/2021
Initial Date Manufacturer Received 01/06/2021
Initial Date FDA Received01/26/2021
Supplement Dates Manufacturer Received05/21/2021
Supplement Dates FDA Received06/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-