• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION TUBE SET, STD VOL, STER; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number H93811
Device Problem Disconnection (1171)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The customer reported it has ¿happened for the past three weeks¿.The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the spikes of an unspecified number of sterile repeater pump tube sets were separated from the tubing; further described as ¿tubing became disconnected at the vial spike and slipping off¿.This issue was identified prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from july 29, 2020 - july 30, 2020.H10: the device was received for evaluation.A visual inspection was performed, and it was noted that the vent filter was detached and missing.No other damage was observed during visual inspection.The spike was received attached to the tubing of the returned sample and through manual pull testing it was noted that the spike could not be detached.The reported condition of a detached spike was not verified; however, it was noted that the vent filter was detached and missing.The cause of the missing vent filter could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBE SET, STD VOL, STER
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11233302
MDR Text Key232158006
Report Number1416980-2021-00188
Device Sequence Number1
Product Code NEP
UDI-Device Identifier00085412475721
UDI-Public(01)00085412475721
Combination Product (y/n)Y
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Type of Report Initial,Followup
Report Date 01/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2023
Device Catalogue NumberH93811
Device Lot Number60249885
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Initial Date Manufacturer Received 12/29/2020
Initial Date FDA Received01/26/2021
Supplement Dates Manufacturer Received02/11/2021
Supplement Dates FDA Received02/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
NA.
-
-