Model Number 71441147
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during inspection it was noticed that the genesis ii mis dcf distal cut blk was returned with a trocar pin stuck in one of the alignment holes.The pin is cold welded and can not be removed.No case reported; therefore, there was no patient involvement.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device has a fractured pin stuck in the pin hole, rendering the device inoperable.The device was manufactured in 2014 and shows signs of extensive use.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII MIS DCF DISTAL CUT BLK
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11235635
MDR Text Key228841133
Report Number1020279-2021-00651
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010497185
UDI-Public03596010497185
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71441147
Device Catalogue Number71441147
Device Lot Number14FM03841
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Date Manufacturer Received02/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-