Catalog Number A2114 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that the knob on top of the mayfield infinity xr2 skull clamp (a2114) that allows it to expand does not hold it in place.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
|
|
Manufacturer Narrative
|
Device history record: the dhr shows no abnormalities related to the reported failure.Mayfield infinity skull clamp (a2114) was returned for evaluation.The reported complaint was confirmed.The left and right pawls are worn and need replaced.Evaluation showed that the torque knob needs to be taken apart and thoroughly cleaned.Unit received without pedi rocker arm or suit case.The unit needs repair, and replacement of worn parts.General maintenance and cleaning required.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
|