OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC PROBE Back to Search Results
Model Number UM-S20-17S
Device Problems Break (1069); Device Contamination with Body Fluid (2317)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/21/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for the evaluation.Omsc found the followings: the visual inspection.There was a hole at the distal part of the sheath of the subject device.Blood was invading from its hole at the distal part.The ultrasonic propagation fluid medium was leaking from its hole at the distal part.There was an indentation on the distal part of the sheath of the subject device.The image inspection.The ultrasound image of the subject device was not drawn normally.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, based on the evaluation, omsc surmised there was the possibility this phenomenon was attributed to occurring a hole at the distal part of the sheath of the subject device due to some external force applied to the distal part by the user handling.Some external force might be swiftly insertion/withdrawing of the subject device from the endoscope with angulation and/or driving the probe.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the user that it was found the dirt on the distal end of the subject device during the unspecified diagnostic procedure with alternate using the subject device and the endoscope.The user tried to wipe the dirt, but the dirt could not be gotten off.After the procedure, the user found the blood inside the distal part of the subject device.There was no patient injury associated with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11242370
MDR Text Key229133808
Report Number8010047-2021-01959
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/26/2020
Initial Date FDA Received01/28/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/19/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-