Model Number UHI-4
Device Problem Gas/Air Leak (2946)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2020
Event Type  malfunction  
Manufacturer Narrative
Since the subject device has been not returned to omsc for the evaluation, the exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the distributor that it was found the co2 gas leakage from the subject device during the installation.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The subject device has not been returned to olympus medical systems corp.(omsc) but was returned to the service department of olympus europe se & co.Kg (oekg) for evaluation.The service department of oekg checked the subject device and confirmed the reported phenomenon.It was found the failure of the first regulator unit and it was required to be replaced this first regulator unit if the subject device was worked.There was no another failure or visible damage.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.However based on the report of the service department of oekg, omsc surmised that the reported phenomenon was attributed to the damage in the internal pressure reducer of the subject device.Since there is no multiplicity, it might have been occurred due to accidental failure.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11243107
MDR Text Key229134725
Report Number8010047-2021-01961
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUHI-4
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-