• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 LONGEVITY HIGH WALL 32MM C; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 LONGEVITY HIGH WALL 32MM C; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Information (3190)
Event Date 01/07/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported during a total hip arthroplasty, that the surgeon stated the acetabular liner wouldn't seat all the way in the acetabular shell.Another liner was used to complete the procedure.Attempts have been made and no further information has been provided.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Complaint sample was evaluated and the reported event was confirmed.A g7 longevity liner was returned.Visual evaluation identified the following: there was damage on the outer spherical surface.All devices that were sent to finished goods inventory were inspected on a cmm at the time of manufacture and were found to be conforming to print specifications.No other damages were noted and no further evaluation could be performed with the returned product.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 LONGEVITY HIGH WALL 32MM C
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11243836
MDR Text Key229138458
Report Number0001822565-2021-00256
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
K190660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 04/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number20123203
Device Lot Number64880210
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/07/2021
Initial Date FDA Received01/28/2021
Supplement Dates Manufacturer Received04/15/2021
Supplement Dates FDA Received04/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-