Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Information (3190)
|
Event Date 01/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported during a total hip arthroplasty, that the surgeon stated the acetabular liner wouldn't seat all the way in the acetabular shell.Another liner was used to complete the procedure.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.A g7 longevity liner was returned.Visual evaluation identified the following: there was damage on the outer spherical surface.All devices that were sent to finished goods inventory were inspected on a cmm at the time of manufacture and were found to be conforming to print specifications.No other damages were noted and no further evaluation could be performed with the returned product.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|