• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Insufficient Information (4580)
Event Date 08/18/2020
Event Type  malfunction  
Manufacturer Narrative
The electrode belt sn (b)(4) was returned and evaluated at the distributor in accordance with zoll manufacturing recommendations.Upon evaluation the electrode belt's shrink tubing was loose from the pulse wire solder joint on the distribution node.The root cause for the separated heat shrink could not be positively identified.No adverse event resulted from the defective belt.
 
Event Description
During an investigation of an electrode belt for an unrelated issue at the distributor, a reportable malfunction was found.The electrode belt's shrink tubing was loose on the pulse wire solder joint in the distribution node.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
matthew saich
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key11246116
MDR Text Key230079004
Report Number3008642652-2021-00796
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005012
UDI-Public00855778005012
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 01/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Device Catalogue Number10A0987
Was Device Available for Evaluation? Yes
Date Manufacturer Received12/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/07/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-