MICROMAR INDUSTRIA E COMERCIO LTDA EASYDRILL PERFORATOR; DRILLS, BURRS, TREPHINES & ACCESSORIES (COMPOUND, POWERED)
|
Back to Search Results |
|
Model Number DM0010FAA |
Device Problems
Device Remains Activated (1525); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
No conclusion can be drawn.No evaluation was performed, as the device was discarded.This dm0010faa easydrill cranial perforator with lot number 715/20 was manufactured by micromar.Multiple warnings are included in the easydrill cranial perforator ifu manual including: it is essential to keep the drill perpendicular (90°) at the predetermined point of the skull to be drilled, as an excessive deviation from perpendicularity may cause the product to fail and lead to serious patient injury.Select a drill bit suitable for the bone thickness.This prevents the drill from tearing the bone or brain tissue (similar effect when the bit is not positioned at90°).If the components of the easydrill cranial perforator become loose during trephination, discontinue use.The drill can cut or tear the dura mater when it unlocks.Check some conditions before performing the procedure, such as the existence of adhering dura mater or other anomalies adjacent to the drilling site.We will continue to track and trend this complaint type.Mhra #:(b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the clutch was not working properly/stuck.It was reported that there was no patient or staff impact.Additional information received, that during a bur hole creation, the clutch of the perforator didn't work which could have caused it to plunge but the surgeon prevent it from happening.It was also reported that there was no procedure delay and it was completed using a backup device.It was also added that the device was discarded.
|
|
Search Alerts/Recalls
|
|
|