Model Number SPL-T
Device Problems Corroded (1131); Incomplete or Inadequate Connection (4037)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and repair.The evaluation confirmed the user report.The unit failed and was pulled out from the generator without having to press on the connector end.In addition, corrosion was found on the metal thread of the connector on the probe side.This event is under investigation.A supplemental report will be submitted upon completion of the investigation.This mdr is being submitted retrospectively as part of a remediation effort related to recent system and process changes.A capa has been opened to manage the actions related to remediation of this issue and any required mdr reporting.
 
Event Description
It was reported that during preparation for use the shockpulse lithotripsy transducer was not working properly.No patient involvement or impact to patient care was reported.
 
Manufacturer Narrative
Upon further review this is not a reportable malfunction.Per the legal manufacturer there is no potential for this issue to cause or contribute to death or serious injury if the malfunction were to recur.Please see the updates in sections: g3, g6, h2, and h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Type of Device
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough PA 01772
MDR Report Key11252060
MDR Text Key250462256
Report Number3011050570-2021-00037
Device Sequence Number1
Product Code FEO
UDI-Device Identifier00821925043831
UDI-Public821925043831
Combination Product (y/n)N
PMA/PMN Number
K142428
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 03/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/09/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/09/2020
Initial Date FDA Received01/29/2021
Supplement Dates Manufacturer Received02/23/2021
Supplement Dates FDA Received03/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-