• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SITE~RITE 8 ULTRASOUND SYSTEM 20 MM PINPOINT GT; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SITE~RITE 8 ULTRASOUND SYSTEM 20 MM PINPOINT GT; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770602Y
Device Problem Poor Quality Image (1408)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/08/2021
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.A history review of serial number (b)(4) showed one similar complaint from this serial number.The similar complaint has been reported by the same us facility.The previous complaint evaluation for this serial number (b)(4) is: inconclusive, no sample returned for evaluation.(reference: mdr number 3006260740-2020-21227).
 
Event Description
Found during evaluation: the probe displays a rain like image.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility with no alleged deficiency reported.During evaluation, it was found probe displays a rain like image.The root cause of the failure was identified as a probe failure.A history review of serial number (b)(6) showed one similar complaint from this serial number.The similar complaint has been reported by the same us facility.The previous complaint evaluation for this serial number (b)(6) is: 1.Inconclusive, no sample returned for evaluation.(reference: mdr number 3006260740-2020-21227) h3 other text : evaluation findings are in section h.11.
 
Event Description
Found during evaluation: the probe displays a rain like image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM 20 MM PINPOINT GT
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
MDR Report Key11252171
MDR Text Key230472468
Report Number3006260740-2021-00155
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741138201
UDI-Public(01)00801741138201
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770602Y
Device Catalogue Number9770602Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/05/2021
Initial Date Manufacturer Received 01/08/2021
Initial Date FDA Received01/29/2021
Supplement Dates Manufacturer Received04/06/2021
Supplement Dates FDA Received04/23/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-