• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS; GAUGE, DEPTH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS; GAUGE, DEPTH Back to Search Results
Model Number 03.130.250
Device Problems Break (1069); Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, during an unknown procedure, it was noticed that the depth gauge was not correct in measuring screws.At some point before the case, the tip of the depth gauge was broken off.No patient harm was done.A different depth gauge was used for the procedure.The procedure was completed successfully.Concomitant devices reported: unknown screws (part#unknown, lot# unknown, quantity unknown) this report involves one (1) depth gauge for 1.3mm/1.5mm and 2.0mm screws.This is report 1 of 1 for pc (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEPTH GAUGE FOR 1.3MM/1.5MM AND 2.0MM SCREWS
Type of Device
GAUGE, DEPTH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK BALSTHAL (CH)
dornacherstrasse 20
balsthal 4710
SZ   4710
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key11252555
MDR Text Key244689845
Report Number2939274-2021-00576
Device Sequence Number1
Product Code HTJ
UDI-Device Identifier10886982075420
UDI-Public(01)10886982075420
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.130.250
Device Catalogue Number03.130.250
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/08/2021
Initial Date FDA Received01/29/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age42 YR
-
-