• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - AMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - AMER Back to Search Results
Model Number 27003
Device Problems Device Displays Incorrect Message (2591); Battery Problem (2885)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2020
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Review of the device data logs confirmed the reported complaint and revealed an error message (sf180) related to a battery charger fault.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to an isolated component failure within the device battery assembly.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an internal battery degraded warning alarm.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - AMER
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AU  2153
MDR Report Key11255471
MDR Text Key229579320
Report Number3007573469-2021-00210
Device Sequence Number1
Product Code NOU
UDI-Device Identifier00619498270033
UDI-Public(01)00619498270033(11)170731(10)1255371
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27003
Device Catalogue Number27003
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/29/2021
Distributor Facility Aware Date01/08/2021
Device Age45 MO
Date Report to Manufacturer01/29/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/29/2021
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-