SMITH & NEPHEW, INC. INTERTAN 1.5 13X40 130D RT; NAIL, FIXATION, BONE Back to Search Results
Model Number 71676626
Device Problems Connection Problem (2900); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that, during a rt hip im nailing, surgeon was unable to fit the lag drill sleeve into the 130 drop.Post procedure, it was noticed there was an issue with the drop being warped.The procedure was completed without delay using a smith & nephew back-up device.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to surgical technique used or user/procedural variance.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERTAN 1.5 13X40 130D RT
Type of Device
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11256939
MDR Text Key229574862
Report Number1020279-2021-00808
Device Sequence Number1
Product Code JDS
UDI-Device Identifier00885556039663
UDI-Public00885556039663
Combination Product (y/n)N
PMA/PMN Number
K040212
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 08/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71676626
Device Catalogue Number71676626
Device Lot Number19BM14853A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/11/2021
Initial Date FDA Received02/01/2021
Supplement Dates Manufacturer Received08/23/2021
Supplement Dates FDA Received08/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-