• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ARCHITECT I1000SR; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ARCHITECT I1000SR; ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE Back to Search Results
Model Number 1L86-01
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/15/2021
Event Type  malfunction  
Manufacturer Narrative
Field service replaced the fitting,quick disc,threaded part on the architect for resolution.A review of the architect serial i1sr51347 service history, revealed no additional issues of splashing from parts reported on the architect serial i1sr51347.No non-conformances or potential non-conformances were identified for the fitting,quick disc,threaded part.A review of the product labeling concluded that the issue is sufficiently addressed.The operations manual also addresses biological and chemical safety hazards that includes wearing personal protective equipment (ppe) and safety awareness where hazards may exist.The i1000sr service and support manual provides instructions for the removal and replacement of the fitting,quick disc,threaded part.No adverse trend was identified for the fitting,quick disc,threaded part.A review of the architect product monitoring found no adverse trend of the architect i1000sr with regards to the complaint issue.No product deficiency was identified for the complaint issue.
 
Event Description
The operator was splashed on intact skin above the neck when the fitting on the accumulator cap broke on the architect i1000sr.Error 5304, (vacuum system error: liquid level too high) was generated on the architect i1000sr.The operator disassembled the rear panel and observed the vacuum accumulator was half full of liquid.As soon as the operator touched the tank, one of the fittings on the cap broke and some liquid splashed which came into contact with intact skin above the neck of the operator.The operator immediately rinsed the area.No injury or medical intervention was reported.No specific patient information provided.No patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT I1000SR
Type of Device
ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Manufacturer (Section D)
ABBOTT LABORATORIES
1915 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT LABORATORIES
1915 hurd drive
irving TX 75038
Manufacturer Contact
christian lee
100 abbott park road
dept. 09b9, lc cp01-3
abbott park, IL 60064-3537
2246682940
MDR Report Key11258065
MDR Text Key229601825
Report Number3016438761-2021-00034
Device Sequence Number1
Product Code JJE
UDI-Device Identifier00380740001537
UDI-Public00380740001537
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1L86-01
Device Catalogue Number01L86-01
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/15/2021
Initial Date FDA Received02/01/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-