Catalog Number A1059
Device Problems Loose or Intermittent Connection (1371); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
This is 2 of 2 reports which is linked to mfg report number 3004608878-2021-00046.A facility reported that the spring came out of the pressure bolt of the mayfield skull clamp (a1059) and the locking handle was loose.It is unknown if there was patient involvement or if the device failure led to patient injury or surgical delay.
 
Manufacturer Narrative
Updated fields: d4, d9, g3, g6, h2, h3, h4, h6, h10 device history record: the dhr shows no abnormalities related to the reported failure.Mayfield modified skull clamp (a1059) was returned for evaluation.The reported complaint was confirmed.Evaluation showed that the device received with the lock has movement and a residue buildup was present.The unit needs repair, and replacement of worn parts.General maintenance and cleaning required.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key11264028
MDR Text Key233500705
Report Number3004608878-2021-00047
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 03/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number169
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/13/2021
Date Manufacturer Received02/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-