MEDTRONIC POWERED SURGICAL SOLUTIONS IPC HANDPIECE ATTACHMENT - INDIGO DRILL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE
|
Back to Search Results |
|
Model Number 1845020 |
Device Problems
Break (1069); Overheating of Device (1437)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis found broken nose bearing and faded laser markings.It was not possible to place a burr in the attachment due to a broken nose bearing, therefore it was not possible to perform the pre-repair heat test.All mandatory parts (including the nose bearing) have been replaced.The laser markings have been redone.The attachment has been successfully tested according to specifications.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
A healthcare provider (hcp) reported that the attachment broke intra-op.On follow-up, it was reported that during drilling, the attachment overheated.The surgeon and the or staff decided to check the angled attachment.After disconnection of the angled attachment some of the balls from the ball bearing came out, so that a further usage of the angled attachment was not possible.A back up device was used to complete the procedure.There was no patient impact.
|
|
Manufacturer Narrative
|
Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: 1845000.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
On follow-up, it was reported that the hand piece overheated.The delay on procedure was the time when the second hand piece being brought to the operating room.
|
|