ADVANCED BIONICS, LLC HIRES ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-04
Device Problems Mechanical Problem (1384); Therapeutic or Diagnostic Output Failure (3023); Appropriate Term/Code Not Available (3191)
Patient Problem Therapeutic Response, Decreased (2271)
Event Date 01/20/2021
Event Type  malfunction  
Event Description
A recipient reportedly experienced decreased performance and sound quality issues.Programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The device passed the external visual and the photographic imaging inspections.System lock was verified.The device passed the electrical tests performed.Advanced electrical testing performed on the device showed lower than typical range on some of the electrodes.It is the opinion of advanced bionics explant review board that some of these electrode test results should be considered as failures.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key11268201
MDR Text Key230985289
Report Number3006556115-2021-00053
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016841002
UDI-Public(01)07630016841002(11)171025(17)201031
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2020
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/02/2021
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age54 YR
-
-