• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SILK ROAD MEDICAL INC. ENROUTE TRANSCAROTID STENT SYSTEM; ENROUTE SDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SILK ROAD MEDICAL INC. ENROUTE TRANSCAROTID STENT SYSTEM; ENROUTE SDS Back to Search Results
Model Number SR-0940-CS
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Brain Injury (2219); Thrombosis/Thrombus (4440)
Event Date 01/19/2021
Event Type  Injury  
Manufacturer Narrative
The product associated with this complaint was not returned to the manufacturer for analysis.A review of manufacturing records for this device was completed and no issues were identified that could have led to the adverse event reported.Although there is no indication that a malfunction of the silk road medical device occurred, the cause of the post-operative complication is unknown.A follow-up mdr will be submitted if additional information is obtained.Silk road medical will continue to monitor for occurrences of similar events.
 
Event Description
It was reported that after the completion of a transcarotid artery revascularization (tcar) procedure, the patient was unable to move her left side.A diagnostic ultrasound (dus) scan was performed that showed a thrombosed stent.The surgeon performed a carotid endarterectomy (cea) and removed the stent to resolve the issue.No further information has been made available by the hospital at this time.At this time there was no post stent imaging made available to silk road medical, no p2y12 inhibitor test/study was performed on the patient, and no further information has been made available by the hospital.The patient was suspected to be covid-19 negative.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE TRANSCAROTID STENT SYSTEM
Type of Device
ENROUTE SDS
Manufacturer (Section D)
SILK ROAD MEDICAL INC.
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
SILK ROAD MEDICAL INC.
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer Contact
magaly boardman
1213 innsbruck drive
sunnyvale, CA 94089
4087309002
MDR Report Key11270625
MDR Text Key233293432
Report Number3014526664-2021-00017
Device Sequence Number1
Product Code NIM
UDI-Device Identifier00811311020508
UDI-Public(01)00811311020508(17)221231(10)301517
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model NumberSR-0940-CS
Device Catalogue NumberSR-0940-CS
Device Lot Number301517
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/31/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Disability;
Patient Age87 YR
-
-