• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXTOME CUTTING BALLOON; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXTOME CUTTING BALLOON; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 3822
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2021
Event Type  malfunction  
Event Description
It was reported that the balloon had a hole.A contralateral approach was used to access the 70% stenosed target lesion in the mildly calcified and mildly tortuous vessel.A 10x2.50 flextome cutting balloon was selected for use.During the procedure inside the patient, the balloon was unable to reach nominal pressure due to a leak from a hole in the balloon.Three attempts were made to inflate the balloon and the balloon was unable to be inflated past 2 atmospheres.The balloon was deflated and removed from the patient.Another 10x2.50 flextome cutting balloon was used to successfully complete the procedure.Here were no patient complications reported and the patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME CUTTING BALLOON
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key11274369
MDR Text Key230243242
Report Number2134265-2021-01175
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeHU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 02/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/05/2023
Device Model Number3822
Device Catalogue Number3822
Device Lot Number0026307453
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/21/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/05/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age74 YR
-
-