• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL 150 - FUKUDA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL 150 - FUKUDA Back to Search Results
Model Number 27093
Device Problem Battery Problem (2885)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/24/2020
Event Type  malfunction  
Manufacturer Narrative
Resmed has requested for the device to be returned so that an engineering investigation could be performed.The device has not been returned, therefore, resmed is unable to confirm the alleged malfunction at this time.(b)(4).
 
Event Description
It was reported to resmed that an astral device displayed a battery communications lost alarm.There was no patient harm or serious injury reported as a result of this incident.
 
Event Description
It was reported to resmed that an astral device displayed a battery communications lost alarm.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Performance testing and review of the device data logs could not reproduce or confirm a battery communications lost alarm.The internal battery was replaced as a precaution.The investigation determined that there was no fault found with the returned device.The device was performing to specifications.Resmed¿s risk associated with use of the device remains acceptable.Resmed reference#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 150 - FUKUDA
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AU  2153
MDR Report Key11276583
MDR Text Key230227728
Report Number3007573469-2021-00226
Device Sequence Number1
Product Code NOU
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Type of Report Initial,Followup
Report Date 03/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27093
Device Catalogue Number27093
Device Lot Number1431751
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2021
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/26/2021
Distributor Facility Aware Date03/15/2021
Device Age11 MO
Date Report to Manufacturer03/26/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/03/2021
Supplement Dates Manufacturer Received03/15/2021
Supplement Dates FDA Received03/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-