Model Number 256045 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medical device lot #: 0093792 was reported, however, this is not a lot# manufactured for this product.Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.(b)(4).
|
|
Event Description
|
It was reported while testing with bd kit flu a+b 30 test physician veritor¿ 5 false positive results were obtained on asymptomatic patients.Confirmatory testing was performed using pcr test method and the results were negative.There was no report of patient impact.
|
|
Manufacturer Narrative
|
H6: investigation summary: this statement is to summarize the investigation results regarding the complaint that alleges false positive results when using kit flu a+b 30 test physician veritor (mn# 256045), batch number 0093792.Bd quality performs a systematic approach to investigate false positive complaints.This approach involves review of manufacturing batch history records, testing of retention samples, and testing of customer returned samples, if applicable.An investigation and testing were performed on the batch number provided.The reported issue was unable to be confirmed.The root cause could not be identified.Bd quality will continue to closely monitor for trends.H3 other text : see h10.
|
|
Event Description
|
It was reported while testing with bd kit flu a+b 30 test physician veritor¿ 5 false positive results were obtained on asymptomatic patients.Confirmatory testing was performed using pcr test method and the results were negative.There was no report of patient impact.
|
|