• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770552
Device Problems Poor Quality Image (1408); No Apparent Adverse Event (3189)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.Device not returned for evaluation.
 
Event Description
The unit experienced a noticeable drop in image quality.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of poor image quality was unconfirmed.The site rite 8 was visually examined upon receipt and was found to be free of physical damage.The reported issue was unconfirmed as the display of the equipment presents good reproduction of the image.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Event Description
The unit experienced a noticeable drop in image quality.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
MDR Report Key11280982
MDR Text Key231411355
Report Number3006260740-2021-00221
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098093
UDI-Public(01)00801741098093
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9770552
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2021
Date Manufacturer Received06/15/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-