• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. LEAD MODEL 302

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. LEAD MODEL 302 Back to Search Results
Model Number 302-20
Device Problems Corroded (1131); Mechanical Problem (1384)
Patient Problem Fibrosis (3167)
Event Date 01/15/2021
Event Type  malfunction  
Event Description
It was reported that the patient was going to have a prophylactic generator replacement with battery.It was indicated that during the battery replacement, the insulation came off from the connector pin.Clarification was provided that the lead body that was coiled behind the generator had scar tissue built up which was due to the body¿s natural response and when the surgeon went in to remove the generator that was being explanted, after pulling it the insulation around the connector pin exposing the wires.That is when the surgeon decided to proceed with a revision.The lead was received into product analysis.No additional relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
dana sprague
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key11281203
MDR Text Key230449223
Report Number1644487-2021-00159
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/02/2007
Device Model Number302-20
Device Lot Number601863
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2021
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 01/15/2021
Initial Date FDA Received02/04/2021
Supplement Dates Manufacturer Received03/31/2021
Supplement Dates FDA Received04/23/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/02/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age20 YR
-
-