OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Residue After Decontamination (2325)
Patient Problem No Patient Involvement (2645)
Event Date 11/16/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation has been completed.A device history record (dhr) review was performed and found no non-conformances that would cause the reported malfunction.All records indicate that the device was manufactured in accordance with all applicable procedures.The instructions for use (ifu) states: ¿the water hardness should not exceed 400 ppm.The recommended water hardness is 150 ppm or lower.¿ an olympus field service engineer (fse) checked the device for residue and no malfunction was found.The root cause could not be conclusively determined.Probable causes that could have led to the reported event include that the hardness of water, which was provided to the subject device, was above 400ppm.Therefore, substance in tap water may have emerged as residue.This mdr is being submitted retrospectively as part of a remediation effort related to recent system and process changes.A capa has been opened to manage the actions related to remediation of this issue and any required mdr reporting.
 
Event Description
A user facility reported that scopes reprocessed in the oer-pro endoscope reprocessor had a white residue on the scopes after reprocessing.There was no patient involvement or injuries reported.No additional information has been obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11281258
MDR Text Key243690142
Report Number8010047-2021-02241
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 02/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/17/2020
Initial Date FDA Received02/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-