• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD. BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM Back to Search Results
Catalog Number 256082
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/14/2021
Event Type  malfunction  
Manufacturer Narrative
A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.(b)(4).
 
Event Description
It was reported while testing for sars cov-2 13 false negative results were obtained.Repeat tests were performed using pcr and the results were positive.There was no indication that results were reported out and there was no report of patient impact.Eua#: (b)(4).
 
Event Description
It was reported while testing for sars cov-2 13 false negative results were obtained.Repeat tests were performed using pcr and the results were positive.There was no indication that results were reported out and there was no report of patient impact.Eua#: (b)(4).
 
Manufacturer Narrative
H.6.Investigation: this statement summarizes the investigation results regarding the complaints that alleges false negative or discrepant results when using kit bd veritor for rapid detection of sars-cov-2 (material # 256082), batch number 0252805.Bd quality performs a systematic approach to investigate false negative complaints.This approach involves review of manufacturing batch history records, testing of retention samples, and testing of customer returned samples if applicable.Investigation and testing were performed on the batch number provided.The complaint was unable to be confirmed.Bd quality will continue to monitor for trends.There were no corrective actions taken at this time.H3 other text : see h.10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD RAPID DETECTION OF SARS-COV-2 VERITOR
Type of Device
CORONAVIRUS ANTIGEN DETECTION SYSTEM
Manufacturer (Section D)
BD RAPIS DIAGNOSTICS (SUZHOU) CO. LTD.
9 rui pu road export zone b
suzhou
CH 
MDR Report Key11282543
MDR Text Key230746672
Report Number3006948883-2021-00189
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 10/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date02/19/2021
Device Catalogue Number256082
Device Lot Number0252805
Was Device Available for Evaluation? No
Date Manufacturer Received10/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-