Model Number 9770501
Device Problems Defective Component (2292); Battery Problem (2885)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2021
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will be evaluated.A supplemental will be submitted with evaluation results.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per tm - battery shutting down at 25-30%.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of battery shutting down at 25-30% was confirmed.The scanner abruptly shuts off before 0% when not connected to ac power.The root cause of the failure was identified as an internal failure within the lithium ion battery.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.H3 other text : evaluation findings are in section h.11.
 
Event Description
Per tm - battery shutting down at 25-30%.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
MDR Report Key11282940
MDR Text Key231407875
Report Number3006260740-2021-00236
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Remedial Action Notification
Type of Report Initial,Followup
Report Date 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2021
Date Manufacturer Received05/04/2021
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction Number2243072-10/11/2019-013-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-