This event has been recorded by zimmer biomet under (b)(4).H6 - investigation conclusions: cause linked to device but unable to trace more specifically.Review of the most recent repair record determined the unit would not power on as the switch was damaged, the insulation on the cable was damaged, and the motor was corroded; however, the repair was not completed because the unit was scrapped locally.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|