An olympus technical support engineer helped the customer to troubleshoot the reported issue.During the troubleshooting, the customer was instructed to stop the process and perform an air purge to drain the basin of the fluid.Per the customer, the lcg usage displayed 2 washes and 8 days, indicating that the acecide-c was expired.The customer conducted a drain and load lcg to load the new acecide-c, high level disinfectant, into the oer-pro.Following these steps, the error code was no longer displayed.If additional information is obtained a supplemental report will be filed.
|
The supplemental report is submitted to provide the result of the legal manufacturer¿s investigation.The device history record for the subject device was reviewed and it was determined that no anomalies were found that could have caused or contribute to the reported problem.The legal manufacturer performed an investigation and confirmed that the user did not replace the disinfectant solution properly which caused the error displayed by the device.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states: prior to reprocessing, check the concentration of the disinfectant solution using a test strip to verify that the concentration of disinfectant solution meets the minimum recommended concentration.Replace the disinfectant solution when it fails to meet minimum recommended concentration or beyond the specified use life.
|