• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE LOW FLOW SYSTEM; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. LEVEL 1 HOTLINE LOW FLOW SYSTEM; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-390
Device Problem Power Problem (3010)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
One level 1 hotline low flow system was received with a cracked tank cover.There was wear and tear damage on the enclosure, line cord, block assembly and front cover.The printed circuit board (pcb) and power switch were outdated.The line cord was plugged into a power source, but there was no power.The front cover and pcb were removed to inspect the power switch to the pcb connection.The reported issue was able to be duplicated.The power switch to pcb connection was damaged due to daily use turning the power on and off.No action was taken to repair the device due to the age and condition of the device.
 
Event Description
Information was received indicating that a smiths medical level 1 hotline low flow system had no power.There was no reported adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEM
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11304240
MDR Text Key231109717
Report Number3012307300-2021-01089
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCON-HL-390
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/13/2021
Initial Date FDA Received02/09/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-