One level 1 hotline low flow system was received with a cracked tank cover.There was wear and tear damage on the enclosure, line cord, block assembly and front cover.The printed circuit board (pcb) and power switch were outdated.The line cord was plugged into a power source, but there was no power.The front cover and pcb were removed to inspect the power switch to the pcb connection.The reported issue was able to be duplicated.The power switch to pcb connection was damaged due to daily use turning the power on and off.No action was taken to repair the device due to the age and condition of the device.
|