• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION SD ALARIS ETCO2; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SD ALARIS ETCO2; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS Back to Search Results
Model Number 8300
Device Problems Excess Flow or Over-Infusion (1311); Inaccurate Delivery (2339)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device has not been received.A follow up report will be submitted with failure investigation results should the device be received for evaluation.
 
Event Description
It was reported there was an infusion error.No additional information.No patient involvement.
 
Manufacturer Narrative
The device has been received and an evaluation is pending.A follow up report will be submitted once the evaluation is completed.
 
Event Description
It was reported there was an infusion error.No additional information.No patient involvement.
 
Event Description
It was reported there was an infusion error.No additional information.No patient involvement.
 
Manufacturer Narrative
A review of the complaint history record in trackwise and sap was performed for the sn (b)(6) which did not confirm similar complaints with the same or related failure mode for this customer.Based on the findings, service determined that the probable root cause of the reported issue was due to electrical failure of the oridion board communication failure.A review of the device history record showed the device had a manufacture date of 09jul2019.The review was performed from the date of manufacture to the date of product release for distribution.A review of the device history record in sap for sn (b)(6) was performed which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS ETCO2
Type of Device
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key11305064
MDR Text Key231159539
Report Number2016493-2021-25264
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10885403830013
UDI-Public10885403830013
Combination Product (y/n)N
PMA/PMN Number
K031741
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 01/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8300
Device Catalogue Number8300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/09/2021
Supplement Dates Manufacturer Received03/23/2021
05/27/2021
Supplement Dates FDA Received04/16/2021
06/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2722-2020
Patient Sequence Number1
-
-