• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBVIE - MEDICAL DEVICE CENTER DUODOPA_DUOPA; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBVIE - MEDICAL DEVICE CENTER DUODOPA_DUOPA; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Catalog Number 062941
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Code Available (3191)
Event Date 08/01/2020
Event Type  Injury  
Manufacturer Narrative
Reference record (b)(4).Catalog number is the international list number which is similar to us list number of 062910.The device involved in the event was not returned remained implanted in the patient was discarded; therefore, a return sample evaluation is unable to be performed.(b)(4).A buried bumper is a known complication of a peg tube placement.If any further relevant information is identified or obtained, a supplemental medwatch will be filed.
 
Event Description
On an unknown date, a patient in (b)(6) underwent a procedure for the placement of percutaneous endoscopic gastrostomy (peg) tube with jejunal (peg-j) tube.In (b)(6) 2020, the patient underwent an endoscopy, which showed ulceration and active inflammation at the internal bumper site.The patient was diagnosed with a buried bumper, but only the j-tube was removed during the endoscopy.The peg tube was left in place.
 
Event Description
Additional information received on 01 feb 2022: on (b)(6) 2021, the the patient underwent a surgical intervention to fix the buried bumper.
 
Manufacturer Narrative
Reference number # (b)(4).The device involved in the event was discarded; therefore, a return sample evaluation is unable to be performed.If any further relevant information is identified or obtained, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUODOPA_DUOPA
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
ABBVIE - MEDICAL DEVICE CENTER
1675 lakeside drive
waukegan IL 60085
Manufacturer (Section G)
ABBVIE - MEDICAL DEVICE CENTER
1675 lakeside drive
waukegan IL 60085
Manufacturer Contact
terry ingram
1675 lakeside drive
waukegan, IL 60085
8479385350
MDR Report Key11308948
MDR Text Key231200525
Report Number3010757606-2021-00094
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K142793
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial,Followup
Report Date 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue Number062941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/21/2021
Initial Date FDA Received02/10/2021
Supplement Dates Manufacturer Received02/01/2022
Supplement Dates FDA Received03/01/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
J-TUBE - UNKNOWN MANUFACTURER AND LOT #; J-TUBE - UNKNOWN MANUFACTURER AND LOT #
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient SexFemale
Patient Weight95 KG
-
-