• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION SUPERION INDIRECT DECOMPRESSION SYSTEM; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE Back to Search Results
Model Number 101-9812
Device Problems Defective Device (2588); Migration (4003)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2021
Event Type  Injury  
Event Description
It was reported that the patient felt the spinous process spacer migrate out of position while he was doing physical therapy.The patient underwent a revision procedure where the spacer was replaced.
 
Event Description
It was reported that the patient felt the spinous process spacer migrate out of position while he was doing physical therapy.The patient underwent a revision procedure where the spacer was replaced.
 
Manufacturer Narrative
The returned spacer was analyzed and revealed severely stripped thread on the actuator.This damage prevented performing the functional test.The damage suggests that the user likely exerted excessive force while attempting to remove the implant, therefore, it is believed that the damage is not related to a device malfunction or to device migration.Per the instructions for use, device migration is a known inherent risk with use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERION INDIRECT DECOMPRESSION SYSTEM
Type of Device
PROSTHESIS, SPINOUS PROCESS SPACER/PLATE
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
suite 100
valencia CA 91355
MDR Report Key11311822
MDR Text Key231304551
Report Number3006630150-2021-00414
Device Sequence Number1
Product Code NQO
Combination Product (y/n)N
PMA/PMN Number
P140004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 03/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number101-9812
Device Catalogue Number101-9812
Device Lot Number800255
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2021
Initial Date Manufacturer Received 01/19/2021
Initial Date FDA Received02/10/2021
Supplement Dates Manufacturer Received02/15/2021
Supplement Dates FDA Received03/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-