Model Number 71440391
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2021
Event Type  malfunction  
Event Description
It was reported that, during a tka procedure, the patella resection guide came loose on instrument.Instruments were outside the patient.There was no delay and procedure was finished with a back up device.
 
Manufacturer Narrative
H6: the device, intended for use in treatment was returned for evaluation: a visual inspection was conducted and confirmed the patella resection guide is damaged.The instrument appears to be become loose when it is on the instrument which causes it to not properly function.A review of complaint history revealed additional complaints for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATELLA RESECTION GUIDE
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11312298
MDR Text Key231302543
Report Number1020279-2021-01123
Device Sequence Number1
Product Code HRY
UDI-Device Identifier03596010453945
UDI-Public03596010453945
Combination Product (y/n)N
PMA/PMN Number
K121393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/11/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440391
Device Catalogue Number71440391
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2021
Date Manufacturer Received07/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-