The equipment was not returned to erbe for an evaluation because it didn't malfunction.No anomalies were found in the review of the device history records (dhrs) for the units.In conclusion, there were no erbe equipment problems that would have caused or contributed to the reported event.There is a warning in the erbe esu and apc user manuals addressing this issue as follows: "interference with cardiac pacemakers, internal defibrillators, or other active implants activation of the electrosurgical unit may affect the performance of active implants or damage them.Risk of injury or death for patients! in the case of patients having active implants, consult the manufacturer of the implant or the competent department of your hospital prior to performing surgery.Do not position the neutral electrode near cardiac pacemakers, internal defibrillators, or other active implants." additionally, when the incident was reported, erbe's clinical education manager provided the account with position statements of recommended practices from several societies and authorities (asge, asa, etc.) involving patients with an aicd.To further address the issue, additional in-service work is being planned with the staff at the medical facility.No trends have been identified with this incident.Erbe usa, inc.Is now closing the file on this event.
|
It was reported that a patient incident occurred with the electrosurgical unit (esu/generator).The esu was used with an argon plasma coagulator (apc, part number 10134-000, associated serial number (b)(4)) and an argon plasma coagulation probe.The patient underwent an upper endoscopy and they had to defibrillate the patient.Additionally, it was reported that the patient had an automatic implanted cardiac defibrillator (aicd).Per the provided information the patient was not injured.
|