MEDTRONIC, INC. DUAL CHAMBER TEMPORARY PACEMAKER; PULSE-GENERATOR, DUAL CHAMBER, IMPLANTABLE
|
Back to Search Results |
|
Model Number 5388XF |
Device Problems
Loose or Intermittent Connection (1371); Use of Device Problem (1670)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/11/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product analysis: analysis of the external pulse generator (epg) confirmed the customer comment that no settings were available (due to empty battery).The device shut down after a few seconds.Analysis noted that the device failed heart wire tests and the wire contacts were worn/contaminated.Analysis also noted that the bails and bail attachments were missing, the ring was bent, a screw was missing from the lower case and the battery chamber was difficult to open.The device has reached end of service and was subsequently scrapped.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that there were no settings on the external pulse generator (epg).The epg was returned for service and subsequently tested out-of-specification during manufacturers analysis.There was no patient involvement.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|